Assay
Standard preparation
Dissolve an accurately weighed quantity of
USP Homosalate RS in chloroform, and dilute quantitatively, and stepwise if necessary, with chloroform to obtain a solution having a known concentration of about 5.0 mg per mL.
Assay preparation
Dissolve about 50 mg of Homosalate, accurately weighed, in 10 mL of chloroform, and mix.
Chromatographic system (see Chromatography
621
)
The gas chromatograph is equipped with a flame-ionization detector and a 0.53-mm × 30-m column coated with a 1-µm film of G27. Hydrogen is used as the carrier gas at a flow rate of about 6 mL per minute. The chromatograph is programmed as follows. Initially the temperature of the column is equilibrated at 70

; upon injection the temperature is increased at a rate of 6

per minute to 220

, and is held at 220

for 5 minutes. The injection port temperature is maintained at 270

, and the detector is maintained at 270

. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the relative retention times are about 0.98 for the
cis-isomer and 1.0 for the
trans-isomer; the resolution,
R, between the
cis-isomer and the
trans-isomer is not less than 1.0; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 1 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of C
16H
22O
3 in the portion of Homosalate taken by the formula:
10C(rU / rS),
in which
C is the concentration, in mg per mL, of
USP Homosalate RS in the
Standard preparation; and
rU and
rS are the sums of the peak responses for the
cis- and
trans-isomers obtained from the
Assay preparation and the
Standard preparation, respectively.