C
46H
77NO
17
916.1
(10
E,12
E)-(3
R,4
S,5
S,6
R,8
R,14
S,15
R)-14-[(6-deoxy-2,3-di-
O-methyl-
-
D-allopyranosyl)oxymethyl]-5-[[3,6-dideoxy-4-
O-(2,6-dideoxy-3-
C-methyl-
-
L-ribo-hexopyranosyl)-3-dimethylamino-
-
D-glucopyranosyl]oxy]-6-formylmethyl-3-hydroxy-4,8,12-trimethyl-9-oxoheptadeca-10,12-dien-15-olide.
Tylosin A
[
1401-69-0].
Packaging and storage
Preserve in well-closed containers, protected from light, moisture, and excessive heat.
Labeling
Label it to indicate that it is for use in animals only.
Identification
A:Ultraviolet Absorption 197U.
Acid solution
Transfer about 50 mg of Tylosin, accurately weighed, to a 100-mL volumetric flask, add 10 mL of 2 N hydrochloric acid, dilute with water to volume, and mix. Transfer 5.0 mL of this solution to a second 100-mL volumetric flask, dilute with water to volume, and mix. Absorptivity at 290 nm is 22.5 ± 2.5, calculated on the dried basis.
Alkaline solution
To 10.0 mL of the final Acid solution add 1.0 mL of 2 N sodium hydroxide, and heat on a water bath for 20 minutes. Cool to room temperature: an absorption maximum is observed at about 332 nm.
B:
The retention time of the major peak for tylosin A in the chromatogram of the Test solution corresponds to that in the chromatogram of the Standard solution, as obtained in the test for Content of tylosins.
Loss on drying 731
Dry about 1 g, accurately weighed, in vacuum at a pressure of not more than 5 mm of mercury at 60
for 3 hours: it loses not more than 5% of its weight.
Residue on ignition 281:
not more than 3.0%, the charred residue being moistened with 2 mL of nitric acid and 5 drops of sulfuric acid.
Limit of tyramine
Transfer 100 mg of it to a 25-mL volumetric flask, add 5.0 mL of 0.03 M phosphoric acid, and swirl to dissolve
(Test solution). Transfer 5.0 mL of a solution containing 70 µg of tyramine per mL in 0.03 M phosphoric acid to a 25-mL volumetric flask
(Standard solution). Transfer 5 mL of 0.03 M phosphoric acid to a 25-mL volumetric flask to provide the blank. Concurrently add to each flask 1.0 mL of a mixture of pyridine and 2.0 mL of filtered ninhydrin solution (1 in 25). Cover the flasks lightly with glass or aluminum foil caps, and heat in a water bath at 85
for not less than 20 minutes. Cool rapidly to room temperature, dilute with water to volume, and mix. Promptly determine the absorbances of the solutions from the
Test solution and the
Standard solution at the wavelength of maximum absorbance at about 570 nm, using the solution from the blank to zero the instrument. The absorbance of the solution from the
Test solution is not greater than that of the solution from the
Standard solution (0.35% of tyramine). In a valid test the solution from the
Standard solution exhibits a dark blue color.
Content of tylosins
Mobile phase
Prepare a mixture of filtered 2 M sodium perchlorate, previously adjusted with 1 N hydrochloric acid to a pH of 2.5 ± 0.1, and acetonitrile (60:40). Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard solution
Transfer about 30 mg of
USP Tylosin RS, accurately weighed, to a 100-mL volumetric flask, add 10 mL of methanol, and swirl to dissolve. Dilute with water to volume, and mix.
Test solution
Transfer about 30 mg of Tylosin, accurately weighed, to a 100-mL volumetric flask, add 10 mL of methanol, and swirl to dissolve. Dilute with water to volume, and mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 280-nm detector, a 4.6-mm × 20-cm column that contains 5-µm packing L1. The flow rate is about 0.7 mL per minute. Chromatograph the
Standard solution, and record the peak responses as directed for
Procedure: the resolution,
R, between the tylosin D peak and the tylosin A peak is not less than 2, the tailing factor is not more than 1.5, and the relative standard deviation for replicate injections is not more than 2%.
Procedure
Separately inject equal volumes (about 20 µL) of the
Standard solution and the
Test solution into the chromatograph, record the chromatograms over a period of time 1.5 times the elution time of the main tylosin A peak, and measure the peak areas for all the peaks. The relative retention times are about 0.5 for tylosin C, 0.7 for tylosin B, 0.9 for tylosin D, and 1.0 for tylosin A. Calculate the percentages of tylosin A, tylosin B, tylosin C, and tylosin D in the Tylosin taken by the formula:
100(ri / rs),
in which
ri is the area of the tylosin A peak, the tylosin B peak, tylosin C peak, or the tylosin D peak, as appropriate, in the chromatogram obtained from the
Test solution, and
rs is the sum of the areas of all of the peaks in the chromatogram obtained from the
Test solution: the content of tylosin A is not less than 80% and the sum of the contents of tylosin A, tylosin B, tylosin C, and tylosin D is not less than 95%.
Assay
Proceed as directed for Tylosin under
AntibioticsMicrobial Assays 81. Prepare the
Test Dilution as follows. Transfer about 250 mg of Tylosin, accurately weighed, to a 500-mL volumetric flask, add 50 mL of methanol, and swirl to dissolve. Dilute with
Buffer No. 3 to volume, and mix. Transfer 4.0 mL of this solution to a second 500-mL volumetric flask, dilute with a mixture of
Buffer No. 3 and methanol (1:1), and mix. This solution contains about 4 µg of tylosin per mL.