Assay
Mobile phase
Dissolve 6.0 g of sodium 1-heptanesulfonate in 3000 mL of water, adjust to a pH of 3.0 by dropwise addition of phosphoric acid. Add 1000 mL of acetonitrile, mix, and pass through a filter having a 0.5-µm or finer porosity. Make adjustments if necessary (see 
System Suitability under 
Chromatography 
621
).
 
Standard preparation
Dissolve an accurately weighed portion of 
USP Xylazine Hydrochloride RS quantitatively in 
Mobile phase to obtain a solution having a known concentration of about 1.2 mg per mL. Transfer 5.0 mL of this solution to a 50-mL volumetric flask, dilute with 
Mobile phase to volume, and mix. Pass a portion of this solution through a filter having a 0.5-µm or finer porosity, discarding the first 3 mL of the filtrate. Use the clear filtrate as the 
Standard preparation. This solution contains about 0.12 mg of 
USP Xylazine Hydrochloride RS per mL.
 
Assay preparation
Transfer an accurately measured volume of Injection, equivalent to about 200 mg of xylazine, to a 100-mL volumetric flask, dilute with Mobile phase to volume, and mix. Transfer 5.0 mL of this solution to a second 100-mL volumetric flask, dilute with Mobile phase to volume, and mix. Pass a portion of this solution through a filter having a 0.5-µm or finer porosity, discarding the first 3 mL of the filtrate. Use the clear filtrate as the Assay preparation.
Chromatographic system
 (see 
Chromatography 
621
)The liquid chromatograph is equipped with a 254-nm detector, a 2-mm × 2-cm guard column that contains packing L2, and a 4.6-mm × 25-cm analytical column that contains packing L1 and is maintained at a constant temperature of about 40

. The flow rate is about 1 mL per minute. Chromatograph the 
Assay preparation, and record the peak responses as directed for 
Procedure: the resolution, 
R, between the main xylazine peak and the closest eluting other peak, if any, is not less than 2.5; and the tailing factor for the xylazine peak is not more than 2.0. Chromatograph the 
Standard preparation, and record the peak responses as directed for 
Procedure: the relative standard deviation for replicate injections is not more than 2.0%. 
[NOTEAfter daily use, rinse the column with 100 mL of water and with 100 mL of methanol, and store the column containing methanol.
]
 
Procedure
Separately inject equal volumes (about 20 µL) of the 
Standard preparation and the 
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of xylazine (C
12H
16N
2S) in each mL of the Injection taken by the formula: 
(220.34 / 256.80)(2000C / V)(rU / rS),
in which 220.34 and 256.80 are the molecular weights of xylazine and xylazine hydrochloride, respectively; 
C is the concentration, in mg per mL, of 
USP Xylazine Hydrochloride RS in the 
Standard preparation; V is the volume, in mL, of Injection taken to prepare the 
Assay preparation; and 
rU and 
rS are the areas of the xylazine peak responses in the chromatograms obtained from the 
Assay preparation and the 
Standard preparation, respectively.